Repairing The Pump - Agilent Technologies 1200 Series Reference Manual

Quaternary pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

2 Insert the socket, delivered with the Seal Wash pump assembly, into the
3 Guide the wire of the active seal wash assembly through the hole and screw
4 Guide the wire over the fan and plug the connector onto the mainboard
Figure 20
5 Replace the foam and top cover.
6 Disconnect all capillaries and tubes from the pump head and disconnect the
7 Using a 4-mm hexagonal key stepwise loosen and remove the two pump
8 Place the pump head, on the backside of the plungerhousing, on a flat
1200 Series QP Reference Manual
hole on the z-panel.
it onto the z-panel.
connector P7.
Wire connected to the mainboard.
active inlet valve cable.
head screws and remove the pump head from the pump drive
surface. Loosen the lock screw (two revolutions) and while holding the

Repairing the Pump

115