If your quaternary pump is part of a complete Agilent 1200 Series system, you
can ensure optimum performance by limiting the configuration of the system
stack to the following configuration. This configuration optimizes the system
flow path, ensuring minimum delay volume.
Solvent
cabinet
Vacuum
degasser
Quaternary pump
Autosampler
Column compartment
Detector
Figure 1
18
Recommended Stack Configuration (Front View)
Flow connections in the stack:
Example setup with 0.17mm ID green capillaries
Solvent bottles - degasser:
G1311-60003 (bottle-head assembly,
PTFE-tubings)
Degasser - pump:
G1322-67300 (PTFE-tubings)
Pump - autosampler:
G1312-67305 (SST, green)
Pump purge valve - waste:
5062-2461 (PTFE tubing wide bore, reorder
pack)
Instant Pilot
Autosampler - column compartment:
G1313-87305 (SST, green)
Column compartment -
column:G1316-87300 (SST, green)
Column - detector:
DAD G1315-87311 (SST, coated)
VWD 5062-8522 (PEEK)
Detector - waste:
DAD 0890-1713 (PTFE, wide bore)
VWD 5062-8535 (PEEK)
5062-2463 (corrugated waste tubing, reorder
pack)