Electrical Connections - Agilent Technologies 1200 Series Reference Manual

Quaternary pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

Electrical Connections

• The GPIB connector is used to connect the pump with a computer. The
• The CAN bus is a serial bus with high speed data transfer. The two
• One analog output provides a pressure signal for integrators or data
• The interface board slot is used for external contacts and BCD bottle
• The REMOTE connector may be used in combination with other analytical
• With the appropriate software, the RS-232C connector may be used to
• The power input socket accepts a line voltage of 100–120 or 220–240 volts
Never use cables other than the ones supplied by Agilent Technologies to ensure
WA R N I N G
proper functionality and compliance with safety or EMC regulations.
1200 Series QP Reference Manual
address and control switch module next to the GPIB connector determines
the GPIB address of your pump. The switches are preset to a default
address
(Table 32
on page 224 or
recognized at powercycling the module.
connectors for the CAN bus are used for internal Agilent 1200 Series
module data transfer and synchronization.
handling systems.
number output or LAN connections.
instruments from Agilent Technologies if you want to use features such as
start, stop, common shut down, prepare, and so on.
control the module from a computer through a RS-232C connection. This
connector is activated and can be configured with the configuration switch
next to the GPIB connector (see
Communication"
on page 231). See your software documentation for
further information.
AC ± 10% with a line frequency of 50 or 60 Hz. Maximum power
consumption is 220 VA. There is no voltage selector on your module because
the power supply has wide-ranging capability. There are no externally
accessible fuses, because automatic electronic fuses are implemented in the
power supply. The security lever at the power input socket prevents the
module cover from being taken off when line power is still connected.
Introduction to the Quaternary Pump
Table 36
on page 229). This address is
"Communication Settings for RS-232C
7
209

Advertisement

Table of Contents
loading

Table of Contents