Pressure sensor calibration wrong or missing

Error ID: 22443
Pressure sensor calibration wrong or missing.
• Parameter: None
Probable cause
1

Seal wash pump was missing when tried to turn on

Error ID: 22499
The seal wash pump has not been detected (while being configured or detected
before)
Probable cause
1
2
3
1290 Infinity II High Speed Pump User Manual
Pressure sensor calibration wrong or
missing.
Defect cable connection to seal wash
pump.
Defect seal wash pump motor.
Defective main board.
Error Information
Pump Error Messages
Suggested actions
• Replace pressure sensor.
• Please contact your Agilent service
representative.
Suggested actions
Check cable connection.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.