Repairing the Pump
Entering the Type Command using the Control Module G1323B
1 Connect the control module to the pump.
2 From Views (press F5) select the System screen, then press Tests (F3).
3 Using the up/down arrows, make sure that the quaternary pump is
4 While in the Tests screen, press m.m (m dot m). From the box now
5 Into the box labeled Instr, enter the command
6 Letters and numbers are created using the up and down arrows. There must
It is important to enter the TYPE command correctly. An incorrect type command may
N O T E
cause the module to turn on in it's resident mode. In such a case, re-enter the TYPE
command correctly.
7 When the command is entered, press Enter to highlight the complete
8 Press the Execute (F8). Below the box, a reply line should then say:
9 Turn off the module, then turn it on again. Turn on should be normal. In the
Entering the Type Command Using the Instant Pilot G4208A
If you are using the Instant Pilot G4208A you will enter the product number of
the quaternary pump (G1310A) without the TYPE command. To learn how to
set the product number, please refer to section
ProductNumber/Serial Number Using the Instant Pilot G4208A"
134
highlighted and press Enter. The Tests screen for the quaternary pump
should now be displayed.
displayed, select the Command line, and press Enter.
TYPE G1311A.
be a blank space between the word TYPE and the letter G.
command.
Reply RA 0000 TYPE "G1311A" .
Records screen, the product# column should indicate the quaternary pump.
If a chemstation is also connected, re-boot it now.
"Entering the
on page 136.