Installing the Pump
3 At the rear of the quaternary pump move the security lever to its maximum
4 Connect the power cable to the power connector at the rear of the
5 Connect the required interface cables to the quaternary pump.
In an Agilent 1200 Series system, the individual modules are connected through CAN
N O T E
cables. The Agilent 1200 Series vacuum degasser is an exception . The vacuum degasser
can be connected via the APG remote connector to the other modules of the stack. The
AUX output allows the user to monitor the vacuum level in the degasser chamber. An
Agilent 1200 Series control module can be connected to the CAN bus at any of the modules
in the system except for the degasser. The Agilent ChemStation can be connected to the
system through one GPIB or LAN (requires the installation of a LAN- board) cable at any of
the modules (except for the degasser), preferably at the detector (MUST for the DAD). For
more information about connecting the handheld controller or Agilent ChemStation refer to
the respective manual. For connecting the Agilent 1200 Series equipment to non-Agilent
1200 Series equipment, see
Slot for interface
board
Analog pressure, 2mV/bar
APG Remote
RS-232C
Figure 4
22
right position.
quaternary pump. The security lever will prevent that the cover is opened
while the power cord is connected to the quaternary pump.
Rear of Quaternary Pump
Chapter
7, "Introduction to the Quaternary Pump"
CAN
Security lever
GPIB
Power
Configuration switch
1200 Series QP Reference Manual