Entering the TYPE Command using the ChemStation
The TYPE is entered by typing a specific command into the command line at
the bottom of the main user interface screen.
1 To enter the TYPE for a specific module, type the following command into
It is important to enter the TYPE command correctly. An incorrect type command may
N O T E
cause the module to turn on in it's resident mode. In such a case, re-enter the TYPE
command correctly.
2 The reply line will respond with RA 0000 Type "G1311A".
3 Turn the quaternary pump off, then on again. Then, re-boot the
4 The TYPE of a module can also be identified by typing the following
Entering the Serial Number using the Control Module G1323B
1 Connect the control module to the quaternary pump. Turn on the pump.
2 In the control module, press Views (F5) and select the
3 Press FW Update (F5). Now, press m. This will display a box which says
4 Press Enter. This will display the box labeled Serial#.
5 Letters and numbers are created using the up and down arrows. Into the
1200 Series QP Reference Manual
the command line:
print sendmodule$(lpmp, "TYPE G1311A")
ChemStation. Boot up and subsequent control of the system should be
normal.
command into the command line:
print sendmodule$ (lpmp, "TYPE?")
The reply line will give the module TYPE.
press Records (F4). Using the up/down arrows, make sure that the
quaternary pump is highlighted.
Update Enter Serial#.
box labeled Serial#, enter the 10-character serial number for the quaternary
pump. When the 10-character serial number is entered, press Enter to
highlight the complete serial number. Then, press Done (F6).