6 Place a new frit into the purge valve with the orientation of the frit as
7 Reinstall the cap with the gold seal.
Before reinstallation always check the gold seal. A deformed seal should be exchanged.
N O T E
8 Insert the purge valve into the pump head and locate the pump outlet
9 Tighten the purge valve and reconnect outlet capillary and waste tubing.
Figure 18
1200 Series QP Reference Manual
shown above.
capillary and the waste tube as shown in
Exchanging the Purge Valve
Repairing the Pump
Figure
18.
Outlet capillary
Purge valve
Waste tube