4
Repairing the Pump
3 Disconnect the two damper capillaries.
5 Disconnect the damper connector at the main board (J24).
Lift the damper out of its position.
Damper
140
4 Loosen the screws of the Z-panel and take it out of the
instrument.
Damper
capillaries
6 Place the new damper into the recess and connect to the
main board (J24).
Damper
Connector
Damper
Damper
Connector
1200 Series QP Reference Manual
Screws
Z-panel