Never operate your instrumentation from a power outlet that has no ground
WA R N I N G
connection. Never use a power cord other than the power cord designed for your
region.
Never use cables other than the ones supplied by Agilent Technologies to ensure
WA R N I N G
proper functionality and compliance with safety or EMC regulations.
Bench Space
The quaternary pump dimensions and weight (see
place the quaternary pump on almost any laboratory bench. It needs an
additional 2.5 cm (1.0 inches) of space on either side and approximately 8 cm
(3.1 inches) in the rear for the circulation of air and electric connections.
If the bench should carry a complete Agilent 1200 Series system, make sure
that the bench is designed to carry the weight of all the modules.
The pump should be operated in a horizontal position!
N O T E
Environment
Your quaternary pump will work within specifications at ambient
temperatures and relative humidity as described in
Do not store, ship or use your quaternary pump under conditions where temperature
C A U T I O N
fluctuations could cause condensation within the quaternary pump. Condensation will
damage the system electronics. If your quaternary pump was shipped in cold weather,
leave it in its box and allow it to warm slowly to room temperature to avoid
condensation.
1200 Series QP Reference Manual
Installing the Pump
Table 1
on page 14) allow to
Table 1
on page 14.