7
Error Information
Pump Error Messages
Pump drive stop not found
Error ID: 22442
The maximum stroke is too long.
• Parameter: 1 – 4 referring to pump drive
Probable cause
1
2
Pressure sensor calibration wrong or missing
Error ID: 22443
Pressure sensor calibration wrong or missing.
• Parameter: None
Probable cause
1
120
Wiper shifted
Pump drive spindle is defect.
Pressure sensor calibration wrong or
missing.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
• Replace pressure sensor.
• Please contact your Agilent service
representative.
Agilent 1290 Infinity Binary Pump User Manual