Agilent Technologies 1200 Series Reference Manual page 136

Quaternary pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

4
Repairing the Pump
For firmware revisions below A02.00 it is very important never to press Done if the Serial#
N O T E
box is blank. In this case, the module can no longer be recognized by either the control
module or the ChemStation. The main board must then be replaced.
6 Turn the quaternary pump off, then on again. The Records screen should
7 If a ChemStation is also connected, re-boot the ChemStation now as well.
Entering the ProductNumber/Serial Number Using the Instant Pilot G4208A
1 Connect the Instant Pilot to the pump's CAN connector. Turn ON the pump.
2 On the Instant Pilot's Welcome screen, press More, then select Maintenance.
3 Press PN/SN. This will display a screen where you can enter the product
4 Make your changes, using the information from the product label of your
To configure the pump as G1311A requires the correct main board version.
N O T E
If you enter the wrong type, your module might not be accessible anymore with the
WA R N I N G
Agilent ChemStation and the Instant Pilot (unsupported module). In such a case
follow the
5 Press OK to highlight the complete command.
6 Press Done to transfer the information into the main board's memory. Press
7 Turn the pump OFF then ON again. The Maintenance screen should display
8 If an Agilent ChemStation is also connected, restart the
136
display the correct serial number.
Using the up/down arrows, select the quaternary pump whose product
number or serial number needs to be changed.
number and/or serial number.
module.
"Recover Instructions"
Cancel quit the process.
the correct serial number for this module.
Agilent ChemStation now as well.
on page 137.
1200 Series QP Reference Manual