Installing The Seal Wash Option - Agilent Technologies 1200 Series Reference Manual

Quaternary pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

4
Repairing the Pump

Installing the Seal Wash Option

When required
Tools required
Parts required
Preparations for this
procedure
1 By using a screwdriver remove the metal plug in the z-panel.
Figure 19
114
When installing seal wash option
4-mm hexagonal key
Screwdriver Pozidrive #1
Active Seal Wash Option kit (Isocratic or Quaternary pump) G1311-68711
Active Seal Wash Option kit (Binary pump) G1312-68711
Switch off the pump at the main power switch
Remove the front cover
Remove the top cover and foam
Removing the metal plug from the z-panel
1200 Series QP Reference Manual