4
Repairing the Pump
Installing the Seal Wash Option
When required
Tools required
Parts required
Preparations for this
procedure
1 By using a screwdriver remove the metal plug in the z-panel.
Figure 19
114
When installing seal wash option
4-mm hexagonal key
Screwdriver Pozidrive #1
Active Seal Wash Option kit (Isocratic or Quaternary pump) G1311-68711
Active Seal Wash Option kit (Binary pump) G1312-68711
Switch off the pump at the main power switch
Remove the front cover
Remove the top cover and foam
Removing the metal plug from the z-panel