Agilent Technologies 1200 Series Reference Manual page 191

Quaternary pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

START and STOP are connected via diodes to pin 3 of the 3394 connector.
N O T E
Agilent 1200 to 3396A Integrators
Connector
03394-60600
Agilent 1200 to 3396 Series II / 3395A Integrators
Use the cable
the integrator prints START; not ready.
1200 Series QP Reference Manual
Pin
3394
9
NC
3
NC
NC
NC
5,14
1
NC
13, 15
03394-60600
and cut pin #5 on the integrator side. Otherwise
Identifying Cables
Pin
Signal Name
Agilent 1200
1 - White
Digital ground
2 - Brown
Prepare run
3 - Gray
Start
4 - Blue
Shut down
5 - Pink
Not connected
6 - Yellow
Power on
7 - Red
Ready
8 - Green
Stop
9 - Black
Start request
Not connected
6
Active
(TTL)
Low
Low
Low
High
High
Low
Low
191

Advertisement

Table of Contents
loading

Table of Contents