Troubleshooting and Diagnostics
Error Messages
Seal Wash (Safety)
Pressure sensor incorrectly plugged into the main board port
C A U T I O N
You risk permanent damage of the pressure sensor, if the seal wash pump is turned ON
while the pressure sensor is incorrectly connected to the main board port, dedicated
for the seal wash pump. This requires exchange of complete multi assembly.
Seal Wash
This error message is generated, if the pressure sensor is incorectly plugged
into the connector on the main board, that should hold the seal wash motor
cable.
Probable cause
1

Wait for Encoder Position Timed Out

Encoder position was not reached within specified time.
Probable cause
1
2
92
Never turn on the seal wash pump, while the pressure sensor is connected to the
main board port.
Make sure the seal wash cable and the pressure sensor cable are plugged in the
correct positions before turning ON the seal wash pump.
Pressure sensor and seal wash pump
connector were connected to the wrong
position on the main board (identical
connectors allow to plug into the wrong
positions by mistake).
Flow too low.
Pump OFF.
Suggested actions
Switch the connectors of the seal wash pump
cable and the pressure sensor to meet the
correct positions (P23 for seal wash cable, P29
for the pressure sensor cable from the multi
assembly).
Suggested actions
Recalculate flow or wait time, to allow to reach
encoder position before encoder position
time-out.
Turn ON pump.