Pressure Signal Missing
The pressure signal from the damper is missing.
The pressure signal from the damper must be within a specific voltage range.
If the pressure signal is missing, the processor detects a voltage of
approximately -120mV across the damper connector.
Probable Causes
• Damper disconnected.
• Defective damper.
Suggested Actions
1200 Series QP Reference Manual
Ensure the damper is connected correctly to the LPM board.
Exchange the damper.
Troubleshooting and Test Functions
3
63