Maintenance

Replace the Seal Wash Pump

Replace the Seal Wash Pump

When
In case of wear of the seal wash pump
Parts required
p/n
5065-4445
5065-9978
1
Lift up solvent filters in solvent reservoirs to avoid
leakages.
2
Open the purge valve.
3
Power off the system.
4
Open the doors.
6
Press the clips.
Preparative Binary Pumps User Manual
Description
Peristaltic pump with Pharmed tubing
Tubing, 1 mm i.d., 3 mm o.d., silicone, 5 m
5
Remove the flow connections from and to the seal
wash pump.
7
Pull the pump to the front.
97