Installing The Seal Wash Option - Agilent Technologies 1200 Series Manual

Quaternary pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

7
Maintenance
Simple Repairs

Installing the Seal Wash Option

When
When installing seal wash option
Tools required
• 4-mm hexagonal key
• Screwdriver Pozidrive #1
Parts required
#
1
1
Preparations
• Switch off the pump at the main power switch
• Remove the front cover
• Remove the top cover and foam
1 By using a screwdriver remove the metal plug in the z-panel.
2 Insert the socket, delivered with the Seal Wash pump assembly, into the
3 Guide the wire of the active seal wash assembly through the hole and screw
4 Guide the wire over the fan and plug the connector onto the mainboard
94
Part number
Description
G1311-68711
Active Seal Wash Option kit (Isocratic or Quaternary pump)
G1312-68711
Active Seal Wash Option kit (Binary pump)
Figure 19
Removing the metal plug from the z-panel
hole on the z-panel.
it onto the z-panel.
connector P7.