4.11 Information on the use of a mobile device with the cockpit app
CAUTION
Improper use of the mobile device
Falling due to altered damping behaviour as a result of unexpected switching into a MyMode.
► Use the instructions for use (user) to instruct the patient on the proper handling of the mobile device with the
Cockpit app.
CAUTION
Independently applied changes or modifications made to the mobile device
Falling due to altered damping behaviour as a result of unexpected switching to a MyMode.
► Do not make any independent changes to the hardware of the mobile device on which the app is installed.
► Do not make any independent changes to the software/firmware of the mobile device that are not included in
the update function of the software/firmware.
CAUTION
Improper mode switching with the mobile device
Falling due to unexpected product behaviour because of changed damping behaviour.
► Ensure that the patient stands securely during all switching processes.
► Inform the patient that the changed damping characteristics have to be verified after switching, and feedback
from the acoustic signal emitter and the mobile device display must be observed.
► Switching back to basic mode is mandatory once the activities in the MyMode have been completed.
NOTICE
Failure to observe the system requirements for the installation of the Cockpit app
Mobile device malfunction.
► The Cockpit App should only be installed on mobile devices and versions which comply with the specifica
tions in the respective online stores (e.g. Apple App Store, Google Play Store, ...)
5 Scope of Delivery and Accessories
5.1 Scope of delivery
•
1 pc. 3C88-3 C-Leg (with threaded connector) or 3C98-3 C-Leg (with pyramid)
•
1 pc. 757L16-4 power supply
•
1 pc. 4E50 Battery Charger for C-Leg
•
1 pc. 4H95 8° C-Leg flexion stop (already installed on delivery)
•
1 pc. cosmetic case for battery charger and power supply
•
1 pc. 646C107 Bluetooth PIN card
•
1 pc. prosthesis passport
•
1 pc. Instructions for use (qualified personnel)
•
1 pc. Instructions for use (user)
•
"4X441-V2=* Cockpit" app for download from the website: https://www.ottobock.com/cockpitapp
5.2 Accessories
The following components are not included in the scope of delivery and may be ordered separately:
•
4H105 knee extender for bench alignment (see page 27)
•
4H106 16° C-Leg flexion stop
•
3S26 cosmetic foam cover
•
3F1=1 Functional cosmesis C-Leg
•
99B120=* Functional stocking
•
4X860=* C-Leg Protective Cover (w/o shield)
•
4P862 C-Leg guard
•
4P863* Shield Insert
•
4X156-1 Charger Extension Cable – Ankle
C-Leg 3C98-3/3C88-3
Scope of Delivery and Accessories
23