CAUTION
Safety mode cannot be activated due to malfunction caused by water penetration or
mechanical damage
Falling due to unexpected behaviour of the product because of changed damping behaviour.
Do not continue using the defective product.
Consult the O&P professional promptly.
CAUTION
Safety mode cannot be deactivated
Falling due to unexpected behaviour of the product because of changed damping behaviour.
If safety mode cannot be deactivated by recharging the battery, a permanent error has
occurred.
Do not continue using the defective product.
The product must be inspected by an authorised Ottobock Service Center. The O&P profes­
sional is your contact.
CAUTION
Safety signal occurs (ongoing vibration)
Falling due to unexpected behaviour of the product because of changed damping behaviour.
Observe the warnings/error signals (see page 95).
Do not continue using the product after the safety signal has been emitted.
The product must be inspected by an authorised Ottobock Service Center. The O&P profes­
sional is your contact.

4.9 Instructions for use with an osseointegrated implant system

WARNING
High mechanical loads due to normal or unusual situations, such as falling
>
Overloading of the bone, which can lead to pain, loosening of the implant, death of bone tis­
sue or bone fracture, among other things.
>
Damage or breakage of the implant system or its components (safety components...).
Verify compliance with the fields of application, conditions of use and indications according
to the information of the manufacturers, both for the knee joint and for the implant system.
Note the instructions of the clinical personnel that indicated the use of the osseointegrated
implant system.
Note changes in your state of health that result in restrictions or doubt regarding the use of
the osseointegrated connection.

4.10 Information on the use of a mobile device with the cockpit app

CAUTION
Improper use of the mobile device
Falling due to changed damping behaviour as a result of unexpected switching to a MyMode.
Make sure you have been instructed in the proper use of the mobile device with the Cockpit
app.
CAUTION
Improper use of the setting parameters in the MyModes
Falling due to unexpected behaviour of the product because of changed damping behaviour.