Setting The Delay Time For Pressure Tolerance Range Alarm; Setting The Pressure Tolerance Range; State Of Alarm - Binder VDL Series Operating Manual

Vacuum drying oven
Hide thumbs Also See for VDL Series:
Table of Contents

Advertisement

With the arrow-down button you can now change to setting the delay for pressure tolerance range
alarm.
With the Back button you can go back to the "Various" submenu and, repeatedly pressing it, to
display.

15.2.3 Setting the delay time for pressure tolerance range alarm

Path:
Normal display
Press the OK button to enable the setting.
Press. alarm delay(min
With the arrow-up button you can go back to the temperature tolerance range setting.
With the arrow-down button you can now change to the pressure tolerance range setting.
With the Back button you can go back to the "Various" submenu and, repeatedly pressing it, to
display.

15.2.4 Setting the pressure tolerance range

Path:
Normal display
Press the OK button to enable the setting
Pressure range(mbar)
With the arrow-up button you can go back to setting the delay for pressure tolerance range alarm.
With the Back button you can go back to the "Various" submenu and, repeatedly pressing it, to
display.

15.2.5 State of alarm

If there are any actual values outside the tolerance range, after the configured alarm delay time the alarm
message "Temperature range alarm" and / or "Pressure range alarm" is displayed in Normal display
(chap. 16.1.1). If the alarm buzzer is activated (chap. 16.4.1) there is an audible alert.
VDL (E3) 09/2019
Settings
Various
Setting the delay for pressure tolerance range alarm.
The current setting flashes. Use the arrow buttons to enter the desi-
red time after which the range alarm shall be triggered. Entry range: 1
up to 120 minutes.
Confirm the entry with the OK button.
Settings
Various
Setting the pressure tolerance range
The current setting flashes. Enter the desired pressure range with the
arrow buttons. Entry range: 10 mbar up to 200 mbar.
Confirm the entry with the OK button.
Normal
Press. alarm delay(min)
Normal
Pressure range(mbar)
Normal
Page 145/234

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vdl 23Vdl 23-ulVdl 56Vdl 56-ulVdl 115Vdl 115-ul

Table of Contents