Safety controller
Note: When the safety controller class 2 had been activated you should disconnect the chamber from the
power supply and have an expert examine and rectify the cause of the fault.
14.2.4 Function check
Check the safety controller at appropriate intervals for its functionality. It is recommended that the autho-
rized operating personnel should perform such a check, e.g., before starting a longer work procedure.
15.
Tolerance range settings
In this menu you can define for temperature and pressure the deviation between the actual value and
setpoint, which that shall cause a tolerance range alarm. The entered value defines the limit of permitted
deviations from the set-point (exceeding and falling below). Reaching this limit triggers tolerance alarm.
In addition you can specify delay times for these alarms.
This function only activates after the set-point has been reached once.
15.1 MB2 controller
15.1.1 Setting the alarm delay times and the tolerance ranges
Main menu > Settings > Various
Path:
• Select the field "Temp. alarm delay" and enter the time in minutes, after which the temperature range
alarm shall be triggered. Setting range: 1 min to 120 min. Confirm entry with Confirm icon.
• Select the field "Temperature range" and enter the desired value for the temperature range. Setting
range: 1,0 °C to 10,0 °C. Confirm entry with Confirm icon.
• Select the field "Press. alarm delay" and enter the time in minutes, after which the pressure range
alarm shall be triggered. Setting range: 1 min to 120 min. Confirm entry with Confirm icon.
• Select the field "Pressure range" and enter the desired value for the pressure range. Setting range:
10 mbar to 200 mbar. Confirm entry with Confirm icon.
VDL (E3) 09/2019
Normal display showing safety controller alarm (sample values)
Submenu "Various".