Display On The Rd4 Controller; Display On The Mb2 Controller; Figure 28: Measuring Connection (12) With Measuring Access Port - Binder VDL Series Operating Manual

Vacuum drying oven
Hide thumbs Also See for VDL Series:
Table of Contents

Advertisement

Technical data of the Pt 100 sensor:
• Three-wire technique
• Class B (DIN EN 60751)
• Temperature range up to 320 °C / 608 °F
• Stainless steel protective tube, length 45 mm / 1.77 in, stainless steel material no. 1.4501

25.4.2 Display on the RD4 controller

In Normal display, the current actual values and the measures object temperature values are shown al-
ternately:
T / Pressure (mbar)

25.4.3 Display on the MB2 controller

The object temperature data are put out together with the data of the temperature controller to the cham-
ber's interface and can be documented by the APT-COM™ 4 Multi Management Software (option, chap.
25.1) developed by BINDER.
VDL (E3) 09/2019

Figure 28: Measuring connection (12) with measuring access port

Obj. Temp.
Normal display with object temperature display
(sample values)
Page 200/234

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vdl 23Vdl 23-ulVdl 56Vdl 56-ulVdl 115Vdl 115-ul

Table of Contents