15.

Tolerance range settings

In this menu you can define for temperature and pressure the deviation between the actual value and
setpoint, which that shall cause a tolerance range alarm. The entered value defines the limit of permitted
deviations from the set-point (exceeding and falling below). Reaching this limit triggers tolerance alarm.
In addition you can specify delay times for these alarms.
This function only activates after the set-point has been reached once.
15.1

Setting the alarm delay times and the tolerance ranges

Main menu > Settings > Various
Path:
• Select the field "Temp. alarm delay" and enter the time in minutes, after which the temperature range
alarm shall be triggered. Setting range: 1 min to 120 min. Confirm entry with Confirm icon.
• Select the field "Temperature range" and enter the desired value for the temperature range. Setting
range: 1,0 °C to 10,0 °C. Confirm entry with Confirm icon.
• Select the field "Press. alarm delay" and enter the time in minutes, after which the pressure range
alarm shall be triggered. Setting range: 1 min to 120 min. Confirm entry with Confirm icon.
• Select the field "Pressure range" and enter the desired value for the pressure range. Setting range:
10 mbar to 200 mbar. Confirm entry with Confirm icon.
For testing the safety temperature limiter (TB) see chap. 16.5.
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press
the Close icon to exit the menu without taking over the entries.

15.1.1 Alarm condition

If there are actual values outside the tolerance range the following information icons for the corresponding
parameter are displayed:
Icon
Signification
"Temperature range"
"Pressure range"
If the condition persists, an alarm is triggered after the configured interval ("range alarm delay"). It is visu-
ally indicated in Normal display. If the alarm buzzer is activated (chap. 16.4) there is an audible alert. The
alarm is shown in the list of active alarms (chap. 16.3).
VDL (E3.1) 06/2020
Submenu "Various".
Information
The current actual temperature value is outside the tolerance range
The current actual pressure value is outside the tolerance range
Page 129/206