The functions are displayed from right to left.
Activated function: switching status "1" (On)
Deactivated function: switching status "0" (Off)
Example:
Activated function "Standby" = 000000000000000
Deactivated function"Standby" = 000000000000000

18.7.4 Setpoint entry

• Select the field "Temperature" and enter the desired temperature setpoint.
Setting range: 0.0 °C up to 220.0 °C
Confirm entry with Confirm icon. The controller returns to the section view.
• Select the field "Pressure" and enter the desired pressure setpoint.
Setting range: 0 mbar up to 1100 mbar
Confirm entry with Confirm icon. The controller returns to the section view.

18.7.5 Tolerance range

You can specify a temperature and pressure program tolerance range for each program section with dif-
ferent values for the tolerance minimum and maximum. When the actual value exceeds the given
threshold, the program is interrupted. This is indicated on the display (see below). When the actual tem-
perature is situated again within the entered tolerance limits, the program automatically continues. There-
fore, the duration of the program may be extended due to the programming of tolerances.
Programming of tolerances may extend program duration.
An entry of "-99999" for the tolerance minimum means "minus infinite" and an entry of "999999" for the
tolerance maximum means "plus infinite". Entry of these values will never lead to program interruption.
The entry of "0" for the tolerance minimum and/or maximum deactivates the respective tolerance function.
When requesting rapid value transitions, we recommend not programming tolerance values in order to
enable the maximum heating-up speed.
VDL (E3) 09/2019
Section view indicating the controller functions.