10.

Set-point entry

Setting ranges
Temperature
0.0 °C / 32 °F up to 220.0 °C / 428 °F
Pressure
0 mbar up to 1100 mbar
After starting the drying procedure, the heater is activated only when the pressure reaches or
falls below the required pressure threshold of 100 mba ist.
Solvents and resulting vapors can ignite at excessive drying temperatures.

10.1 MB2 controller

In Fixed value operating mode you can enter a temperature set-point, a pressure set-point, and the swit-
ching-state of special controller functions.
All settings made in Fixed value operating mode remain valid until the next manual change. They are
saved also when turning off the chamber or in case of toggling to Program Mode.
The values entered in Fixed value mode become valid again after the end of a program and
will then be targeted and equilibrated.
When operating without a vacuum by setting "Pressure control off" (chap. 11.5.1), the pres-
sure tolerance range function is automatically deactivated.

10.1.1 Set-point entry through the "Setpoints" menu

Press the Setpoint setting icon to access the "Setpoint" setting menu from Normal display.
VDL (E3) 09/2019
Control ranges
10 °C / 18 °F to 16 °C / 28.8 °F above ambient
temperature up to 110 ºC / 230 °F
0 mbar up to 1100 mbar
"Setpoints" menu.
Select "Fixed value operation setpoints" to access the
individual parameters.