After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press
the Close icon to exit the menu without taking over the entries.
15.1.2 Alarm condition
If there are actual values outside the tolerance range the following information icons for the corresponding
parameter are displayed:
Icon
Signification
"Temperature range"
"Pressure range"
If the condition persists, an alarm is triggered after the configured interval ("range alarm delay"). It is visu-
ally indicated in Normal display. If the alarm buzzer is activated (chap. 16.4) there is an audible alert. The
alarm is shown in the list of active alarms (chap. 16.3).
15.2 RD4 controller
Required access level: "Admin".
15.2.1 Setting the delay time for temperature tolerance range alarm
Path:
Normal display
Press the OK button to enable the setting.
Temp. alarm delay(min)
With the arrow-down button you can now change to the temperature tolerance range setting.
With the Back button you can go back to the "Various" submenu and, repeatedly pressing it, to
display.
15.2.2 Setting the temperature tolerance range
Path:
Normal display
Press the OK button to enable the setting.
Temperature range(°C)
With the arrow-up button you can go back to setting the delay for temperature tolerance range alarm.
VDL (E3) 09/2019
Information
The current actual temperature value is outside the tolerance range
The current actual pressure value is outside the tolerance range
Settings
Various
Setting the delay for temperature tolerance range alarm.
The current setting flashes. Use the arrow buttons to enter the desi-
red time after which the range alarm shall be triggered. Entry range: 1
up to 120 minutes.
Confirm the entry with the OK button.
Settings
Various
Setting the temperature tolerance range
The current setting flashes. Enter the desired temperature range with
the arrow buttons. Entry range: 1.0 °C up to 10.0 °C
Confirm the entry with the OK button.
Temp. alarm delay(min)
Temperature range(°C)
Page 144/234
Normal