25.
Options
25.1
APT-COM™ 4 Multi Management software (option)
The chamber is regularly equipped with an Ethernet interface (2) that can connect the BINDER APT-
COM™ 4 Multi Management Software.
• RD4 controller: The MAC Address is indicated in the "Ethernet" controller menu (chap. 20.2.1.1).
• MB2 controller: MAC Address is indicated in the "Device info" controller menu (chap. 20.1.1.2).
The actual temperature and pressure values are given at adjustable intervals. Programming can be per-
formed graphically via PC. Up to 100 chambers can be cross-linked. For further information please refer
to the APT-COM™ 4 operating manual.
25.2 Analog outputs for temperature and pressure (option)
With this option, the camber is equipped with analog outputs 4-20 mA for temperature and pressure.
These outputs allow transmitting data to external data registration systems or devices.
The connection cable is firmly connected to the connection "Analog output" (3b) located in the VDL rear
control panel. The connection is realized as a 9-poles SUB-D socket as follows:
Figure 26: Pin configuration of the SUB-D socket „Analog output" (8) for the analog outputs option
25.3
Measuring access port for vacuum, 9 poles (option)
The measuring access port allows creating electrical connections for low voltages or sensors between
the exterior and interior of the VDL vacuum drying oven. A 9 poles plug for the outside is included
VDL (E3) 09/2019
ANALOG OUTPUT 4-20 mA DC
PIN 1: Temperature +
PIN 2: Temperature –
PIN 4: Pressure +
PIN 5: Pressure –
Temperature range: 0 °C to +110 °C
Pressure range: 0 mbar to 1100 mbar
A suitable plug is enclosed.
Figure 27: Measuring connection (12) with measuring access
port and supplied plug
Page 198/234