16.2.2 RD4 controller
In the event of equipment failures, when the temperature and / or pressure deviate from the set tolerance
range limits, optical and possibly acoustic alarm messages are given out via the controller.
The alarm messages on leaving the tolerance range are emitted after a configurable delay (chap. 15.2.1,
15.2.2), the others immediately when the fault occurs. The tolerance range alarm is suppressed after
turning on the chamber until the setpoint is reached and then for the selected delay time.
In Normal display a text message indicates the alarm cause. The "collective alarm" icon flashes. If
the audible arm is activated, the buzzer sounds.
If more than one alarm signal is sent simultaneously, they are displayed in a cycle.
Press the OK button to confirm the alarm and mute the buzzer. If the alarm cause is still valid, the "coll-
ective alarm" icon is lit.
Safety controller
Alarm messages overview:
Condition
Tolerance range alarm: actual tem-
perature value exceeding the tole-
rance range)
Tolerance range alarm: actual pres-
sure value exceeding the tolerance
range)
Exceeded setpoint of the safety con-
troller class 2
Overtemperature protective device
class 2 (thermal switch) activated
Safety temperature limiter (TL) has
turend off the heater
VDL (E3) 09/2019
Normal display in state of alarm (example).
(a) Screen header flashing in red color and
(b) Alarm icon on the bottom of the screen:
Alarm indication (example: safety controller alarm)
Alarm message
"Temperature range alarm"
"Pressure range alarm"
"Safety controller"
"Overtemperature"
"Overtemperature"
showing the alarm message
change to the list of active alarms and alarm
acknowledgement
Start after
condition occurred
after configurable time
(chap. 15.2.1)
Factory setting: 10 minutes
after configurable time
(chap. 15.2.2)
Factory setting: 10 minutes
immediately
immediately
immediately
Page 150/234