Menu Structure Of The Controller And Access Levels - Binder VDL Series Operating Manual

Vacuum drying oven
Hide thumbs Also See for VDL Series:
Table of Contents

Advertisement

Icon
Signification
Back button
Standby button

8.2.1 Menu structure of the controller and access levels

Starting from Normal display, navigate between the menus with the arrow buttons. With the OK button
you enter the setting of further subordinate menu functions. With the Back button you go back to the
previous function and finally back to Normal display.
Starting from Normal display you can directly access the quick access menu functions with the OK but-
ton.
The available functions depend on the current dependent on the current authorization "User", "Admin" or
"Service", for which the entry of a password may be required, depending on the setting.
You can set passwords for different access levels:
• User: The password enables access to the standard operating functions. Factory setting is 00 00 (no
password assigned).
• Admin: The password enables access to advanced controller functions and settings. Factory setting
is 00 01.
• Service: The password enables access to all controller functions (for BINDER Service only).
As soon as a password has been assigned, access to the respective functions is blocked and only
available after entering the correct password.
Menu
Required access level
Quick access
Any user
Setpoints
"User"
Programs
"User"
Chamber info
Any user
Settings
"Admin"
VDL (E3) 09/2019
Function
• Back to previous menu level.
• Press down 5 seconds to simultaneously display the inner chamber
temperature and the heating element temperature (chap. 13.2.8)
Turning on 7 off the chamber (chap. 9.5)
Functions
• Turning on / off pressure control and temperature
control
• Starting / stopping Drying monitoring
• Temperature and pressure set-point setting
• Setting the safety controller
• Turning on / off the vacuum pump and the optional
connection "GAS/AIR 2" (10)
• Closing all fine dosing valves
• Start / Stop function for time and week program
• Configuration display (setup information, controller
hardware and software, analog inputs)
• Display of interface configuration (e.g. MAC address,
IP address)
• General controller settings (date, time, menu language,
temperature unit, display brightness...)
• Network settings
• Setting the data logger storage interval
• Setting the tolerance ranger limits and delay times for
tolerance range alarms
• Password changing for User and Admin
Page 88/234

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vdl 23Vdl 23-ulVdl 56Vdl 56-ulVdl 115Vdl 115-ul

Table of Contents