Set-Point Entry; Set-Point Entry Through The "Setpoints" Menu - Binder VDL Series Operating Manual

Vacuum drying oven with microprocessor program controller mb2
Hide thumbs Also See for VDL Series:
Table of Contents

Advertisement

10.

Set-point entry

Setting ranges
Temperature
0.0 °C / 32 °F up to 110 ºC / 230 °F
Pressure
0 mbar up to 1100 mbar
After starting the drying procedure, the heater is activated only when the pressure reaches or
falls below the required pressure threshold of 100 mbar.
Solvents and resulting vapors can ignite at excessive drying temperatures.
Fire and explosion hazard when exceeding the auto-ignition temperature of the sol-
vent due to excess drying temperature.
Serious injury or death from burns and / or explosion pressure.
 Make sure that only solvents with an auto-ignition temperature that is higher than
200 °C / 392 °F may be entered into the chamber. Check the auto-ignition temperature
with the safety specifications of the solvent. In the case of solvent mixtures, use the au-
to-ignition temperature of the material with the lowest auto-ignition temperature.
In Fixed value operating mode you can enter a temperature set-point, a pressure set-point, and the
switching-state of special controller functions.
All settings made in Fixed value operating mode remain valid until the next manual change. They are
saved also when turning off the chamber or in case of toggling to Program Mode.
The values entered in Fixed value mode become valid again after the end of a program and
will then be targeted and equilibrated.
When operating without a vacuum by setting "Pressure control off" (chap. 11.5), the pressure
tolerance range function is automatically deactivated.

10.1.1 Set-point entry through the "Setpoints" menu

Press the Setpoint setting icon to access the "Setpoint" setting menu from Normal display.
VDL (E3.1) 06/2020
Control ranges
10 °C / 18 °F above ambient temperature up to
110 ºC / 230 °F
0 mbar up to 1100 mbar
DANGER
"Setpoints" menu.
Select "Fixed value operation setpoints" to access the
individual parameters.
Page 101/206