Maintenance Intervals - Dräger Evita 2 Instructions For Use Manual

Intensive care ventilator
Table of Contents

Advertisement

Maintenance Intervals
Clean and disinfect
equipment
and/or components
before any maintenance
procedures*-
and before returning
for repair.
02 sensor capsule
Ambient-air filter
Cooling-air filter
Lip seals of
connecting plate on
back of patient system
NiCd battery for
mains power alarm
failure
Lithium battery for
data storage
Pressure reducer
Equipment check and
service
To be replaced by trained service personnel
every 2 years.
,
Disposal, page 102.
To be replaced by trained service personnel
every 2 years.
Disposal, page 102.
Complete overhaul every 6 years by trained
service personnel.
We recommend DrsgerService.
Every six months by trained service personnel.
Regular safety checks
Every six months by trained service personnel.
Replace sensor capsule if display: 02 meas. inop.,
and if calibration is not possible, page 8.
Disposal, page 102.
Replace after 1 year, page 103.
Dispose of in normal way.
Clean or replace after 4 weeks, page 102.
Replace every year.
Dispose of in narmal way.
To be replaced by trained service
personnel every year.
If the glass on the LCD display breaks, a liquid chemical may escape. This
should not be allowed to come into contact with the human body. In
emergencies, clean affected skin with soap.
101