Section 5 - Operation
Open Reservoir Scavenger
Connecting the
Open Reservoir
Scavenger System
5-5-2
The open reservoir scavenger system is installed on the Narkomed 2B
before shipping. The only thing you need to do before operating the
scavenger is to make the hose connections.
CAUTION:
Take special care not to accidentally force 19 mm
scavenger hoses over 22 mm breathing hose terminals.
Carefully follow the hose connection instructions for
installing the scavenger and the absorber.
To connect the scavenger hoses:
1. Attach a 19 mm scavenger hose between the 19 mm terminal
(marked SCAVENGER HOSE) on the bottom of the absorber pole and
the 19 mm terminal (marked SCAVENGER HOSE) on the right side
of the scavenger.
WARNING: Make sure the 19 mm scavenger hoses leading from the
absorber are not pinched, kinked, or blocked in any
manner.
2. Attach the short 19 mm scavenger hose between the 19 mm terminal
(marked SCAVENGER HOSE) on the rear of the APL valve and the
19 mm terminal (marked SCAVENGER HOSE) on the rear of the
absorber pole.
3. Attach another 19 mm scavenger hose between the 19 mm terminal
(marked SCAVENGER HOSE) on the ventilator relief valve and the
19 mm terminal (marked SCAVENGER HOSE) on the left-hand side
of the scavenger.
WARNING: Make sure the 19 mm scavenger hose leading from the
ventilator relief valve is not pinched, kinked, or blocked in
any manner.
4. Attach a wall suction hose between the wall suction outlet and the
suction terminal (DISS or hose barb with adapter) on the scavenger.
5. Verify the proper functioning of the scavenger system.