Slope Trigger - Rigol DS6000 Series User Manual

Digital oscilloscope
Hide thumbs Also See for DS6000 Series:
Table of Contents

Advertisement

RIGOL

Slope Trigger

Trigger on the positive or negative slope with specified time.
Trigger Type:
Press Type to select "Slope". At this point, the trigger setting information as shown
in the figure below is displayed at the upper-right corner of the screen.
Source Selection:
Press Source to select CH1-CH4 as the Trigger Source. The current trigger source
is displayed at the upper-right corner of the screen.
Note: Select channel with signal input as trigger source to obtain stable trigger.
Slope Condition:
Press When to select the desired slope condition.
: trigger when the positive slope time of the input signal is greater than
the specified Time.
: trigger when the positive slope time of the input signal is lower than the
specified Time.
: trigger when the positive slope time of the input signal is greater than
the specified Lower Limit of Time and lower than the specified Upper Limit of
Time.
: trigger when the negative slope time of the input signal is greater than
the specified Time.
: trigger when the negative slope time of the input signal is lower than
the specified Time.
: trigger when the negative slope time of the input signal is greater than
the specified Lower Limit of Time and lower than the specified Upper Limit of
Time.
5-12
5 To Trigger the Oscilloscope
DS6000 User's Guide