Video Standard:
Press Standard to select the desired video standard.
NTSC: the field frequency is 60 fields per second and the frame frequency is 30
frames per second. The TV sweep line is 525 with the even field goes first and
the odd field follows behind.
PAL/SECAM:
PAL: the frame frequency is 25 frames per second. The TV sweep line is 625
with the odd field goes first and the even field follows behind.
SECAM: the frame frequency is 25 frames per second. The sweep line is 625
with interlacing sweep.
480P: the frame frequency is 60 frames per second; the TV scan line is 525;
progressive scan; the line frequency is 31.5 kHz.
576P: the frame frequency is 60 frames per second; the TV scan line is 625;
progressive scan.
HDTV: HDTV consists of 720P, 1080P, and 1080I display formats. The specified
video standards are as follows:
720P: the frame frequency is 60 frames per second; the TV scan line is 750;
progressive scan; the line frequency is 45 kHz.
1080P: the frame frequency is 60 frames per second; the TV scan line is
1125; progressive scan; professional format.
1080I: the field frequency is 50 to 60 fields per second; the frame
frequency is 25 to 30 frames per second; the TV scan line is 1125;
interlaced sweep; the line frequency is 33.75 kHz.
Trigger Mode:
Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto,
normal or single. The corresponding status light of the current trigger mode turns on.
Trigger Level:
Use TRIGGER
description of Trigger Level on page 5-8.
5-16
LEVEL to modify the level. For details, please refer to the
5 To Trigger the Oscilloscope