5 To Trigger the Oscilloscope
RIGOL

Trigger Mode

Trigger mode would influence the mode in which the oscilloscope searches for trigger
event. The following is the schematic diagram of the acquisition memory from which
you can see that the position of the trigger event is determined by the reference time
point and the delay setting.
Trigger Event
Pre-trigger Buffer
Post-trigger Buffer
Acquisition memory
Pre-trigger/ Delayed trigger:
Acquire data before/after the trigger event. The trigger position is usually at the
horizontal center of the screen. In full-screen display, you can view seven-grid
pre-trigger and delayed trigger information. You can adjust the horizontal position of
the waveform through HORIZONTAL
POSITION to view more pre-trigger
information or delayed trigger information, through which to know the signal
information before/after the trigger (e.g. capture the glitch generated by the circuit
and analyze the pre-trigger data to find out the reasons for generating glitch).
Press MODE in the trigger control area (TRIGGER) on the front panel or through the
MENU  Sweep menu to select the desired trigger mode. The corresponding
status light of the mode currently selected turns on.
MODE