RIGOL
SLOPe Trigger
1.
:TRIGger:SLOPe:TIME
Command Format:
:TRIGger:SLOPe:TIME <count>
:TRIGger:SLOPe:TIME?
Function Explanation:
The commands are used to set and query the slope time from 20ns~10s.
Returned Format:
Query returns the set value of <count> in scientific numeric notation in s.
Example:
:TRIG:SLOP:TIME 0.01
:TRIG:SLOP:TIME?
2.
:TRIGger:SLOPe:SENSitivity
Command Format:
:TRIGger:SLOPe:SENSitivity <count>
:TRIGger:SLOPe:SENSitivity?
Function Explanation:
The commands are used to set and query the trigger sensitivity of SLOPe
trigger. The range of <count> is 0.1~1.
Returned Format:
Query returns the set value of <count> in scientific numeric notation in div.
Example:
:TRIG:SLOP:SENS 0.5
:TRIG:SLOP:SENS?
2-30
Set the slope time as 10ms
Query returns 1.000e-02
Set the trigger sensitivity as 0.5div
Query returns 5.00e-01
Programming Guide for DS1000CA Series
Programming Introduction