RIGOL
5. Sample Point
The sample point is the point within the bit time. The oscilloscope samples the
bit level at this point. The sample point position is expressed by the percentage
of the "time from the start of bit to the sample bit" takes up in "bit time".
Press Sample Point and use
and the range is from 5% to 95%.
6. Threshold
Refer to the introduction in "Parallel Decoding".
7. Display-related Setting
Press Format to set the display format of the bus to Hex, Decimal, Binary or
ASCII.
Press Offset and use
Press BusStatus to enable or disable bus display.
8. Decoding Table
The decoding table lists the decoded data, the corresponding line number, time,
error information, and other information in table format. It can be used to
observe relatively longer decoded data to solve the problem that some data
could not be seen clearly on the screen.
7-22
to adjust this parameter with a step of 1%
1 bit
60%
70%
80%
to adjust the vertical display position of the bus.
7 Protocol Decoding