RIGOL
nnn is from 1 to 223 (except 127) and the ranges of the other three nnn are
from 0 to 255. You are recommended to ask your network administrator for an
IP address available.
Press IP Address and use
be saved in the non-volatile memory and if "Power-off Recall" is set to "Last"
and DHCP and Auto IP is set to "Off" at the next power-on, the oscilloscope
will load the preset IP address automatically.
2. Set the Subnet Mask.
The format of the subnet mask is nnn.nnn.nnn.nnn, where, the range of the nnn
is from 0 to 255. You are recommended to ask your network administrator for a
subnet mask available.
Press Mask and use
saved in the non-volatile memory and if "Power-off Recall" is set to "Last" and
DHCP and Auto IP is set to "Off" at the next power-on, the oscilloscope will
load the preset subnet mask automatically.
Set the Gate
You can set this paramter In Auto IP and Static IP mode.
The format of the gate is nnn.nnn.nnn.nnn, where, the range of the first nnn is from
1 to 223 (except 127) and the ranges of the other three nnn are from 0 to 255. You
are recommended to ask your network administrator for a gate address available.
Press Gate and use
in the non-volatile memory and if "Power-off Recall" is set to "Last" and DHCP
and Auto IP is set to "Off" at the next power-on, the oscilloscope will load the preset
gate address automatically.
13-4
to input the desired IP address. This setting will
to input the desired subnet mask. This setting will be
to input the desired gate address. This setting will be saved
13 System Function Setting
DS6000 User's Guide