LED on
LED on
power sup
battery
ply
charger
Error after disconnecting the battery charger (error on self-test)
Beep sig
Error
nal
5 x long
Self-test error
17.2.3 Status signals
Battery charger is connected
LED on
LED on
power sup
battery
ply
charger
Battery charger disconnected
Beep sig
Event
nal
1 x short
Self-test completed successfully.
Battery charge level
The current battery charge level can only be displayed by connecting the battery charger and power supply to the
product.
Battery
charger
Battery charge level less than 50%
Battery charge level over 50%
C-Brace
Error
Defective power supply
No connection between battery charger
and power supply
Defective battery charger
Battery is fully charged (or connection
with product is interrupted).
Event
Power supply and battery charger operational
Resolution
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Check whether the charging cable plug is
fully engaged on the battery charger.
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Take note of the different beep signals to
identify the given charging state.
Upon connecting or disconnecting the
battery charger, a self-test is conducted
which is confirmed by beep signals.
When these beep signals are heard, the
battery is fully charged.
If no beep signals are heard, the connec
tion to the product is interrupted.
If the connection to the product is inter
rupted, an authorised Ottobock Service
Centre must inspect the product, battery
charger and power supply.
Resolution
•
Restart the product by connecting and dis
connecting the battery charger (up to three
times)
•
If the error persists, the product must be
inspected by an authorised Ottobock Service
Centre
Appendices