To switch from a C-Leg fitting to a fitting with this product, close the Ottobock DataStation, restart it and enter
the user name for the fitting with this product.

8.6.2 Establishing a connection between the product and C-Soft

The following points need to be observed for establishing the connection:
The 60X5 BionicLink PC is connected to the PC being used and the corresponding drivers are installed.
The 60X7 BionicLink is connected to the product.
The product battery is charged (the blue LED on the 60X7 BionicLink is lit up or flashing).
The battery charger is disconnected from the product.
The product is within the range of the Bluetooth connection.
The range is around 10 m but can vary depending on local conditions.
After entering the password, the Establish connection tab is displayed automatically.
The product emits beep signals in order to indicate to the patient and orthotist that the connection has been estab­
lished. The blue LED on the BionicLink flashes as well.
INFORMATION
If the connection is disrupted during operation, see the section "Troubleshooting" (see Page 84).
Once the connection has been established, the data are read from the product.
Disconnecting from the product
1) Click the Establish connection tab.
2) Click the Disconnect button.
→ The connection to the product is disconnected.
3) Disconnect the BionicLink from the product.
4) Close the plug cover.
Reconnecting to the product
1) Connect the BionicLink to the product.
2) Click the Establish connection tab.
3) Click the Reconnect button.
4) Read the security notice on the screen and click the OK button.
→ A connection is established to the product.

8.6.3 Calibration (zero setting)

INFORMATION
Calibration is required for all products adjustable with C-Soft.
Adjustments are not possible without prior calibration!
Instruct the patient in accordance with the explanations in the software's adjustment section.
INFORMATION
Calibration is required each time a connection has been established.
Calibration while standing
Once the connection has been established, the Calibration tab is displayed automatically.
C-Brace
1) Click the Connect via Bluetooth button to establish a Bluetooth connection
between C-Soft and the product.
INFORMATION: When several products are present, select the desired
BionicLink device according to the information on the reverse side of the
device. If only one BionicLink device is available, the selection will be
skipped automatically.
2) Read the security notice on the screen and click the OK button.
→ A connection is established to the product.
Adjustment with C-Soft
63

Advertisement

Chapters

Table of Contents
loading