Information On Alignment/Adjustment - Otto Bock C-Brace Instructions For Use Manual

Table of Contents

Advertisement

Available languages

Available languages

Safety
CAUTION
Product not set upright for storage/charging
Falling due to unexpected behaviour of the product as a result of changed damping behaviour caused by air in the
hydraulic unit.
► Always keep the product upright for charging and storage.
► Before using the product, verify that the damping behaviour corresponds to the chosen mode.
► Take note of noises or inconsistent damping behaviour.
► Walk 10 to 20 steps in order to activate the automatic bleeding process.
NOTICE
Use of incorrect power supply unit/battery charger
Damage to product due to incorrect voltage, current or polarity.
► Use only power supply units/battery chargers approved for this product by Ottobock (see instructions for use
and catalogues).

4.6 Information on Alignment/Adjustment

CAUTION
Disconnecting/establishing the connection during the adjustment process with the adjustment software
Falling due to unexpected behaviour of the product.
► When wearing the product, the patient must not remain unattended during the configuration process while
connected to the adjustment software.
► Observe the maximum range of the Bluetooth connection.
► During the data transfer (PC to product), the patient must remain standing or sitting still and the BionicLink or
the communication cable must not be removed.
► If the connection fails while making adjustments, the orthotist must immediately warn and secure the patient.
► The connection to the product must always be disconnected after adjustments have been completed.
CAUTION
Operator errors during the adjustment process with the adjustment software
Falling due to unexpected behaviour of the product.
► Participation in an Ottobock product training course is mandatory prior to using the product. Additional
product training courses may become necessary to qualify for software updates.
► If the patient uses walking aids (e.g. crutches or walking canes) during the adjustment process, readjustment
is required as soon as the patient stops using these walking aids.
CAUTION
Damage to the plug contacts of the BionicLink connection
Falling due to unexpected behaviour of the product.
► After completing the adjustments, ensure that the plug cover on the hydraulic unit is completely closed.
CAUTION
Error during adaptation to the patient
>
Falling due to unexpected behaviour of the product
>
Chafing during relative flexion between product and leg
► During adaptation to the patient, note the correct position of the product knee rotation point (height and A-P
alignment).
► Note that the product is a monocentric knee joint so that slight movements between the frame structure and
leg can occur.
► Avoid wearing the product directly on the skin.
52