Adjustment with C-Soft
The
symbol is displayed after calibration is successfully completed.
If the
symbol is displayed, calibration could not be performed. To correct errors, read the section "Troubleshoot­
ing" (see Page 84). Perform calibration again after correcting errors.
Calibration while sitting
Once the connection has been established, the Calibration tab is displayed automatically.
1) The patient has to sit in a chair of suitable height.
2) Extend the leg with the product, the edge of the heel touches the floor.
INFORMATION: Neither a heel nor a toe load should be shown on the screen in this position.
3) Click the Calibration button.
→ The product is calibrated.
The
symbol is displayed after calibration is successfully completed.
If the
symbol is displayed, calibration could not be performed. To correct errors, read the section "Troubleshoot­
ing" (see Page 84). Perform calibration again after correcting errors.
INFORMATION
Do not use the hands to help extend the leg.

8.6.4 Adjustment procedure on product

INFORMATION
The program does not perform any automatic changes of the adjustments! Each change must be made or con­
firmed by the orthotist!
The adjustment software describes the procedure for each adjustment step in the adjustment section.
► After successful calibration, click the Settings tab.
→ The product adjustments are displayed.

8.6.5 Overview of adjustment parameters

The adjustment parameters of the adjustment software are described in more detail in the section that follows. In
this overview, the sequence of the parameters is based on the movement pattern or the gait phases when walking
on level ground that are relevant for the parameters. The sequence in the description therefore deviates from the
listing in the C-Soft adjustment software.
The denotation of the parameters differentiates between the stance phase (STP = Stance Phase) and the swing
phase (SWP = Swing Phase).
STP Flexion Damping
The parameter 'STP Flexion Damping' is the resistance against flexion of the knee which is required for descend­
ing stairs or ramps or for supported sitting motions.
The value configured on delivery has to be adapted to the patient.
64
1) The patient must stand between parallel bars or with crutches.
2) Extend the leg with the product to the rear with no movement in the knee joint
and no load.
3) Click the Calibration button.
→ The product is calibrated.