Adjustment with C-Soft

8.5 Using the Ottobock DataStation

For the use of the Ottobock DataStation, please read the adjustment software instructions for use. You can also
consult the integrated online help of the adjustment software for further information.

8.6 C-Soft application

CAUTION
Disconnecting/establishing the connection during the adjustment process with the adjustment software
Falling due to unexpected behaviour of the product.
► When wearing the product, the patient must not remain unattended during the configuration process while
connected to the adjustment software.
► Observe the maximum range of the Bluetooth connection.
► During the data transfer (PC to product), the patient must remain standing or sitting still and the BionicLink or
the communication cable must not be removed.
► If the connection fails while making adjustments, the orthotist must immediately warn and secure the patient.
► The connection to the product must always be disconnected after adjustments have been completed.

8.6.1 Starting the Program

C-Soft can be started in the following ways:
New job
1) Click the Add Job button.
→ The tab opens.
2) Click the C-Soft button.
If a lower limb application has already been installed, the button is on the Prosthetic Lower Extremity tab.
Existing job
The job must have been created previously using C-Soft 2.6 or higher.
► Double-click the existing job.
Entering the password
The password input screen is displayed after starting the Ottobock DataStation > C-Soft.
INFORMATION
Unlock PIN: This code is valid on any computer.
Personal password: This code is only valid on the computer on which it was entered/changed.
INFORMATION
If certification for the C-Leg has been obtained, the user name is not valid for the fitting with this product. Only by
entering the user name for the fitting with this product will the C-Soft version for this product be available.
Certification for the C-Leg cannot be used for this product. An error message is therefore displayed when estab­
lishing a connection.
62
Before the first use, the Username and Unlock-PIN fields must be filled in with
the relevant information that was given to you during the Ottobock product train­
ing course for the adjustment software.
Information: make sure to use correct spelling (upper/lower case) when
entering the information.
Click the Log On button.
After entering the unlock PIN and clicking the Log On button you will be asked
to enter and confirm a password of your own choice.
Information: the password entry window only appears after you have
entered the unlock PIN. If you have already entered a password in the pre­
vious step, the password entry screen will not appear.
Click the OK button.