Scope of Delivery and Accessories

4.11 Notes on the safety modes

CAUTION
Safety mode cannot be deactivated
Falling due to unexpected product behaviour because of changed damping behaviour.
► If you cannot deactivate safety mode by recharging the battery, a permanent malfunction has occurred.
► Do not continue using the defective product.
► The product must be inspected by an authorised Ottobock Service Centre.
CAUTION
Safety signal occurs (ongoing beep signal)
Falling due to unexpected product behaviour because of changed damping behaviour.
► Observe the warning/error signals (see Page 82).
► After the safety signal has been emitted, further use of the product is prohibited.
► The product must be inspected by an authorised Ottobock Service Centre.
CAUTION
Using the product in safety mode
Falling due to unexpected product behaviour because of changed damping behaviour.
► The warnings/error signals (see Page 82) have to be observed.
5 Scope of Delivery and Accessories
Scope of Delivery
1 pc. 757L16* power supply
1 pc. 4E50 Battery Charger for C-Leg
1 pc. leg orthosis (aligned) with 17B300=* C-Brace joint unit
1 pc. 647G631 instructions for use (qualified personnel)
1 pc. 646D458 instructions for use (user)
1 pc. 647F449 orthotic passport
1 pc. cosmetic case for XEKT2337 battery charger
Accessories
"C-Soft V2.8" or higher adjustment software
Update from C-Soft 2.0 4X180=V2.0, C-Soft 2.2 4X180=V2.2, C-Soft 2.4 4X180=V2.4, C-Soft 2.6 4X180=V2.6
via Internet download
60X7 BionicLink
60X5 BionicLink PC
INFORMATION
The adjustment software only works correctly with the "60X5 BionicLink PC" Bluetooth adapter. All other
Bluetooth adapters, including those integrated into the computer, are not supported.

6 Charging the battery

The following points must be observed when charging the battery:
The capacity of a fully charged battery is sufficient for one full day.
We recommend charging the product once a day when used by the patient on a daily basis.
When used daily, the complete charging unit (power supply – battery charger) may remain plugged into the
wall socket.
The battery should be charged for at least 3 hours prior to initial use.