Scope of Delivery and Accessories
4.11 Notes on the safety modes
CAUTION
Safety mode cannot be deactivated
Falling due to unexpected product behaviour because of changed damping behaviour.
► If you cannot deactivate safety mode by recharging the battery, a permanent malfunction has occurred.
► Do not continue using the defective product.
► The product must be inspected by an authorised Ottobock Service Centre.
CAUTION
Safety signal occurs (ongoing beep signal)
Falling due to unexpected product behaviour because of changed damping behaviour.
► Observe the warning/error signals (see Page 82).
► After the safety signal has been emitted, further use of the product is prohibited.
► The product must be inspected by an authorised Ottobock Service Centre.
CAUTION
Using the product in safety mode
Falling due to unexpected product behaviour because of changed damping behaviour.
► The warnings/error signals (see Page 82) have to be observed.
5 Scope of Delivery and Accessories
Scope of Delivery
•
1 pc. 757L16* power supply
•
1 pc. 4E50 Battery Charger for C-Leg
•
1 pc. leg orthosis (aligned) with 17B300=* C-Brace joint unit
•
1 pc. 647G631 instructions for use (qualified personnel)
•
1 pc. 646D458 instructions for use (user)
•
1 pc. 647F449 orthotic passport
•
1 pc. cosmetic case for XEKT2337 battery charger
Accessories
•
"C-Soft V2.8" or higher adjustment software
Update from C-Soft 2.0 4X180=V2.0, C-Soft 2.2 4X180=V2.2, C-Soft 2.4 4X180=V2.4, C-Soft 2.6 4X180=V2.6
via Internet download
•
60X7 BionicLink
•
60X5 BionicLink PC
INFORMATION
The adjustment software only works correctly with the "60X5 BionicLink PC" Bluetooth adapter. All other
Bluetooth adapters, including those integrated into the computer, are not supported.
6 Charging the battery
The following points must be observed when charging the battery:
•
The capacity of a fully charged battery is sufficient for one full day.
•
We recommend charging the product once a day when used by the patient on a daily basis.
•
When used daily, the complete charging unit (power supply – battery charger) may remain plugged into the
wall socket.
•
The battery should be charged for at least 3 hours prior to initial use.
58
C-Brace