7. Messages
Selection process aborted due to algorithm revision number change.
The Boot and Algorithm disks are not the same version. Check your disks to make sure that the two
match. Re-select the desired device, and the device algorithm will be loaded after the new device menu
data is loaded.
SELFTEST FAILED:
An error was detected during the self-test.
SELFTEST FAILED: Failure of Mass Storage Module - Drive x.
This message appears if the Mass Storage Module fails the on-line self-test. This error may be caused by a
hardware problem.
Single drive copy overwrites user RAM. Hit Return to continue, ^Z to abort.
If you try to perform a single drive disk copy, the programmer uses user RAM as a buffer. Therefore,
anything already in user RAM will be overwritten when copying to the same drive. If you don't want the
user RAM to change, press C
+ Z.
TRL
System error. Please contact Data I/O.
Contact Data I/O Customer Support.
System Files disk not found.
When the yield tally option is enabled, the System Files disk will be searched. If the disk is not found, this
message appears.
System initialization failed.
An error occurred while the programmer was booting. The problem may be caused by a defective disk, try
another copy of the Boot Files disk.
System parameters restored.
This message appears when you have restored a configuration file from the Restore System Parameters
screen.
System parameters saved.
This message appears when you save a set of system parameters.
Task error.
UniSite-xpi has experienced an error that it cannot recover from; turn UniSite-xpi off and reboot the
system. If the error recurs, contact Data I/O Customer Support.
Terminal port is not properly connected.
This message will appear if the user attempts to perform an operation involving the terminal port when it
is not connected properly. This may occur when the user has terminal functions assigned to the remote
port and an attempt is made to perform a data transfer operation over the terminal port when nothing is
connected to it. This could also be the result of the user's attempting to enter transparent mode with
nothing connected to the terminal port.
TEST ABORTED: Base not installed.
If you selected the Adapter/Relay test from the Self-test screen, and a Base is not installed, this message
will appear. Install a Base and try the self-test again.