Mnemonic
Name
Command
IC
Internal Clock
Rate
48
Quasonix, Inc.
Description
Report or set the internal clock
rate
This rate is used if the clock
source is set to internal (CS 1). It
should not be confused with "BR",
which sets the rate of the internal
bit sync, which phase locks to the
externally applied data, if Clock-
free is enabled.
If no argument is passed, the unit
reports the clock frequency. If a
valid frequency is given, the
internal clock frequency is set.
The frequency is in MHz.
Examples:
IC
Display current internal
clock rate
IC 4.95 Set internal clock rate to
4.95 MHz
Valid range is 0.002 MHz –
46.000 MHz
Observes same bit rate limits as
HR/LR cmds (PCM/FM half)
Refer to section 4.2.3.1.3 for
additional IC command detail
TIMTER™ Multi-mode Dual
Telemetry Transmitter
Option (s)
Setting
Required
Saved?
Standard
Y
Factory
Default
IC 5
Need help?