...11 MAINTENANCE
o) Isolate the katharometer unit at its PSU.
p) Make the remaining electrical connections at TB1 of the
katharometer unit – see Fig. 5.5 on page 17.
q) Replace the cover of the katharometer unit.
r) Power up the katharometer unit from its PSU.
s) Carry out a calibration procedure – see Section 8.3.
Note. It is possible that the zero reading may drift for
several days after the removal of liquid.
11.4.2 Removal/Replacement of a Display Unit
a) Electrically isolate the 6553 Display Unit.
b) Release the retaining screw through the display facia and
carefully withdraw the chassis from its edge connectors and
out through the front panel – see Fig. 3.1 on page 5.
c) To replace the unit, carefully insert it into the display facia and
press firmly into position before tightening the retaining
screw.
d) Power up the display unit and carry out a calibration – see
Section 8.3.

11.4.3 Error Messages

If the error message 'NV Memory Error' is displayed the contents
of the non-volatile memory has not been read correctly during
power up.
To rectify the fault, switch off, wait 10 seconds and switch on
again. If the fault persists, contact the Company.

12 SPARE PARTS LIST

Warning. Interference with any unit or its components
implies acceptance of responsibility by that person for
ensuring the continuing maintenance of intrinsic safety
requirements. Unauthorized repair/ use of spare parts
or incorrect assembly may render any unit unfit for use
in an intrinsically safe application.
Note. Although the digital display units may be marked
4600 on their display facia, they are dedicated variants
which are not interchangeable with the Company's
standard 4600 Controller/Display. These dedicated
display units are identified (4689 502) as shown in Fig.
3.1 on page 5.
When ordering a katharometer unit, it is necessary to specify the
zero gas and range in association with the Company part
number. See the typical identification label shown in Fig. 3.2 on
page 5.
12.1 Consumables
Description
Model 006548 000 & 006548 203
Katharometer Analyzer Panels
Granular anhydrous CaCl
2
12.2 Routine Maintenance Parts
Description
Model 6553 Display Unit
Fuse, 500mA a/s 20 x 5 mm glass cartridge
Function selector switch, 3 position, 2 wafer
Potentiometer (1k ), zer o adjustment
Katharometer Analyzer Panel
Seal, top of drying chamber
Seal, bottom of drying chamber
Gauze, drying chamber
12.3 Repair Maintenance Parts
Description
Model 4234 Power Supply Unit
Nominal 230V unit
Nominal 115V unit
Fuses
F2/F3 – 250mA/ 1500A HBC cartridge
F1 – 400mA cartridge
Katharometer Analyzer Panel
Flowmeter
Valve, metering
Coupling seal ring
Katharometer unit
006548 001
Model 6553 Display Unit
Display units (H
/CO
)
2
2
Display units (H
/Ar)
2
Zener diode safety barrier devices MTL 7055ac
Zener diode safety barrier devices MTL 7755ac
Part No.
Sourced locally
Part No.
0231 538
0234 724
002569 036
006540 203 006548 000
002310 012 002310 012
006519 160
0211 035
006525 700 006548 018
Part No.
4234 500
4234 501
0231577
0231555
006540 203 006548 000
006525 460
0216 485
006525 480
0216 484
006525 130
006539 960K (or J)
4689 502
4689 505
0248 297
0248 296
33

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ak103Ak101Ak104

Table of Contents