Chapter 4: Temperature Measurement
Note: This feature is optional on Root.
For temperature measurements obtained with the TIR-1 bluetooth thermometer, refer to the
TIR-1 manual. Root offers an optional integrated temperature module and takes a
temperature measurement through the use of a temperature probe. The temperature probe is
designed for use on adult and pediatric patients. Patient temperature can be measured via an
oral/axillary or rectal probe.
Operation - Temperature
Spot Check Mode
Spot Check mode provides a one-time predictive measurement that takes a temperature in
approximately 6 to 15 seconds, before a steady state temperature is achieved. Predictive
measurements reduce the time required for measurement by using algorithms to predict
what the temperature would be if the probe were left in place until steady state is achieved.
Continuous Mode
Continuous mode provides continuous temperature readings for up to 10 minutes through a
direct measurement. Direct measurement is used to continuously read temperatures until it
reaches the thermal steady state (unchanging). The thermal steady state is achieved through
an oral or rectal measurement in approximately 3 minutes and through an axillary
measurement in approximately 5 minutes.
Taking Temperature Measurement
1.
Ensure that the correct measurement site is selected before measurement.
Note: If a rectal probe is attached, the Site button
Note: The site selection toggle is disabled once the probe has been placed on the
site. The toggle will be re-enabled once the probe has been returned to the probe
well holder.
2.
To change the measurement site, touch the Main Menu icon
Temperature Settings