Root NIBPT
Procedure
1.
Connect the manometer, 500mL bottle, and hand bulb to the NIBP module using
"T" adapters and connection tubing.
2.
Power ON the manometer.
Set the unit of measurement to mmHg.
Verify the manometer has been calibrated.
3.
Power ON the Root.
4.
On the Root home screen, press the Main Menu icon
Press NIBP
5.
6.
Press CALIBRATION
For the password, enter 4258.
Press the Enter key.
Press AIR LEAK TEST
7.
To begin the air leak test, press Test.
8.
Wait for the countdown timer to reach 0 second.